Table 2.

Frequency of ≥3-category insulin dosing errors as a function of imprecision (percent CV) using the gaussian error model.

Imprecision, % CV≥3-category dosing errors, %
50.00
60.00
70.00
80.00
90.02
100.05
110.12
120.25
130.45
140.74
151.10
161.57
172.12
182.75
193.46
204.22