Table 3.

Concentrations corresponding to 10% CV imprecision and 99th percentile reference limit for the evaluated troponin assays.

Platform99th percentile limit,1 μg/L10% total CV concentration, μg/LRatio of 10% CV concentration to 99th percentile limit
AxSYM0.301.224.1
ACS:1800.100.373.7
Centaur0.100.333.3
Immuno 10.100.343.4
Access0.040.061.5
Access 20.040.092.3
Vidas0.100.363.6
Liaison0.030.0652.2
Dimension RxL0.070.263.7
Opus0.100.909.0
Stratus CS0.070.101.4
Immulite One0.200.321.6
Alpha Dx0.15ND2
Vitros ECi0.100.444.4
E1700.010.044.0
Elecsys 10100.010.044.0
AIA 210.060.091.5